Zero-distance Drop Test System mainly simulates the resistance to drop and impact of large and heavy packaging products. Its strong power system and unique sample holder facilitate easy loading and unloading of oversized and overweight items and automatically rise to the setting. After the height, the drop test is completed. It can realize the drop test of the edge, surface and angle of the sample.
This equipment is mainly used to evaluate the ability of product or packaging to withstand drops during transportation and loading and unloading, so as to improve product and packaging design.
Features:
-
Driven by pneumatic and servo motors, stable lifting process with upper and lower displacement restrictions, safe and reliable;
-
Adopt single-track or dual-track lifting method, and the height can be adjusted arbitrarily;
-
It can clamp and drop the test specimen in different directions such as edges, faces and angles;
-
The testing machine adopts PLC full-automatic control, large size true color touch LCD screen,
-
No special foundation required, no other complicated operation or installation;
-
Remote control can be realized to ensure the safety of operators (customized functions).
Model | SM11-100 | SM11-200 | SM11-300 | SM11-500 | SM11-800 | SM11-1000 | SM12-100 | SM12-200 |
Parameters | ||||||||
Max. Load (kg) |
100 | 200 | 300 | 500 | 800 | 1000 | 100 | 200 |
Drop Height (mm) |
0~1200 | 0~1200 | 0~1200 | 0~1200 | 0~1000 | 0~1000 | 300~1500 | 300~1500 |
Max. Specimen Size (mm) |
1000×1000×1000 | 1200×1200×1200 | 1300×1300×1300 | 1500×1500×1000 | 1600×1600×800 | 1800×1800×600 | 1000×1000×1000 | 1200×1200×1200 |
Single Lift Drop Zone Size(W*D/mm) | 1000×1200 | 1000×1200 | 1000×1200 | 1200×1500 | 1200×1200 | 1400×1400 | ||
Double lift Drop Zone Size(W*D/mm) | 1700×1400 | 1700×1400 | 1900×1600 | 2000×1800 | 2200×1900 | 2300×2000 | ||
Test Mode | Face, Edge and Angle | |||||||
Overall Dimension (mm) |
2000×1200× | 2100×1200× | 2200×1200× | 2500×1500× | 2000×1200× | 2200×1400× | ||
2000 | 2200 | 2300 | 2500 | 2600 | 2800 | |||
Double lift Overall Dimension (mm) |
2050×1400× | 2050×1400× | 2250×1600× | 2350×1800× | 2550×1900× | 2650×2000× | ||
1615 | 1615 | 1615 | 1515 | 1400 | 1300 | |||
Weight (kg) |
1700 | 1800 | 2000 | 2500 | 3000 | 3200 | 1500 | 2000 |
Consumption Power (kVA) |
1.5 | 2 | 3 | 5.5 | 7.5 | 11 | 2 | 3 |
Power Supply for Tester | AC220V±10%,50Hz | |||||||
Power Supply for Air Source |
AC220V±10%,50Hz,3kVA /AC380V±10%, 50Hz,5kVA | |||||||
Air Source Conditions | The output pressure is no greater than 1.0MPa, and air compressor is needed to be configured if there is no air source in the lab. | |||||||
Working Environment |
Temperature range 0~40℃,humidity≤80% (no condense) | |||||||
Standards | ISO2248-1985 (E) IEC68-2-27 |