Model: AX-127B
High Speed・High Accuracy 100kHz
Digital ESR/Z Checker
Most Suitabe Model to inspect and measure ESR/Z of Low Impedance-Capacitor
● Measurement of low-equivalent series resistance and low-impedance on 100kHz at high-stablility・high-accuracy・high-speed.
● Low-ESR/Low-impedance measuring range: 100mΩ~100Ω range.
● Resolution:10μΩ (ESR)
● Available to make the Contact・Check to watch the contact condition of Probe. Contact.
● 4 digit display and LO/GO/HI/ Contact・Error Output.
● By RS-232C or GP-IB, all functions can be controlled externally (Option)
● Printer output (Option)
Specifications
Measuring range and Accuracy (at23℃±5℃) [When the standard setting]
Test Current | Measuring range | Resolution | Measuring current | ESR Accuracy | Z Accuracy |
---|---|---|---|---|---|
100mΩ | 0.00mΩ~109.99mΩ | 0.01mΩ | 100mA | within ±0.5% of rdg±10digit ±100μΩ |
within ±1% of rdg ±10digit ±100μΩ |
1Ω | 0.0mΩ~1099.9mΩ | 0.1mΩ | 10mA | within ±0.5% of rdg ±10digit |
within ±1% of rdg ±10digit |
10Ω | 0.000Ω~10.999Ω | 1mΩ | 1mA | ||
100Ω | 0.00Ω~109.99Ω | 10mΩ | 0.1mA |
※The above accuracy is the vaue was made zero adjustment exactly after 30 minutes of warming-up.
Measuring method | 5-terminal measuring [available to contact check] |
---|---|
Measuring frequency | 100kHz±0.1%, SINE wave |
Measuring time | [Free running] FAST:10 times/sec. approx. SLOW:2.5 times/sec. [When the standard setting] |
[Remote start] [ESR] about (4.2+N) ×2ms [N=FAST:1~99] /SLOW: [4+power line cycle×1~9] ×2ms (please plus 1ms., in case of making the Contact-ckeck) |
|
[Remote start mode] [Z] about (3.5+N) ×3ms [N=FAST:1~99] /SLOW: [3.5+power line cycle×1~9] ×3ms (please plus 1ms., in case of making the Contact-ckeck) |
|
Comparator set range | 9999 both for HI and LO |
Operation condition | [Temp.] 5℃~+50℃ [Humidity] less than 85% |
Power supply | AC100V/117V/220V/240V selectable, 50/60Hz, about 30VA |
Outer dimension | 333 (W) ×99 (H) ×300 (D) mm (excluding protruding parts such as rubber legs, etc.) |
Weight | about 4kg |
Option